A PIND test is a Particle Impact Noise Detection test, performed on high-reliability electronic components to detect loose particles inside of sealed cavities within the device. The test was developed by NASA, McDonnell Douglas, and Texas Instruments before being published under Test Method 2020 in MIL-STD-883. The test provides a nondestructive means of identifying devices containing particles of sufficient mass that upon impact within the cavity, excite the transducer. Particles may be introduced into the device at any point in the manufacturing process up to and including the lid seal operation. Prior to implementation of this test, particle contamination was identified to be responsible for several catastrophic failures of the Delta Launch Vehicle Program.
Common sources of particle contamination include:
The PIND test is specified in the following standards:
These tests are required for certain classes of devices in the generic performance specification standards. MIL-PRF-38535 requires all Class S devices to undergo testing in accordance with MIL-STD-883 TM 2020, condition A. For hybrid devices, MIL-PRF-38534 requires all class K devices to be screened in accordance with MIL-STD-883 TM 2020, conditions A or B. For discrete devices, MIL-PRF-19500 requires testing in accordance with MIL-STD-750 TM 2052, Condition A or B for all JANTX, JANTXV, and JANS devices.
During testing, the device is coupled to a transducer, which is vibrated so that any particles present in the cavity will impact the walls of the device. The output of the transducer is amplified, and fed through a band pass filter to remove system noise and the shaker frequency. This is then fed into a threshold detector, oscilloscope, and speaker. Indication of a particle detected through any of the three outputs constitutes a failure.
The test apparatus consists of the following:
After failure, particles may be removed by the following procedure and evaluated to find the root cause and perform corrective actions. After puncturing the device lid, cover with tape. Vibrate the device until all loose particles are captured in the tape, confirmed by a successful test. The particles can then be examined and characterized by SEM or ESCA.
Empirical verification of the test with devices seeded with particles found that only 40-60% of devices were correctly identified, and found false positives in 10-20% of known clean samples. Accordingly, the test methods in MIL-STD-883 and MIL-STD-750 call for the test to be repeated if failures are detected in >1% of devices, and require a lot failure if >25% of devices fail or the test exceeds 5 rounds.