A colorimeter is a device used in colorimetry that measures the absorbance of particular wavelengths of light by a specific solution. It is commonly used to determine the concentration of a known solute in a given solution by the application of the BeerâÂÂLambert law, which states that the concentration of a solute is proportional to the absorbance.
The essential parts of a colorimeter are:
In addition, there may be:
There are many commercialized colorimeters as well as open source versions with construction documentation for education and for research.
Changeable optics filters are used in the colorimeter to select the wavelength which the solute absorbs the most, in order to maximize accuracy. The usual wavelength range is from 400 to 700 nm. If it is necessary to operate in the ultraviolet range then some modifications to the colorimeter are needed. In modern colorimeters the filament lamp and filters may be replaced by several (light-emitting diode) of different colors.
In a manual colorimeter the cuvettes are inserted and removed by hand. An automated colorimeter (as used in an AutoAnalyzer) is fitted with a flowcell through which solution flows continuously.
The output from a colorimeter may be displayed by an analogue or digital meter and may be shown as transmittance (a linear scale from 0 to 100%) or as absorbance (a logarithmic scale from zero to infinity). The useful range of the absorbance scale is from 0 to 2 but it is desirable to keep within the range 0âÂÂ1, because above 1 the results become unreliable due to scattering of light.
In addition, the output may be sent to a chart recorder, data logger, or computer.
In clinical laboratories, the colorimeter is commonly used to estimate various biochemical compounds inàbiological samples. In all methods where a colored product is formed in reaction with a specific analyte, the analyte can be quantitatively measured. For instance, it is used in the FolinâÂÂWu method for measuring blood glucose, in which glucose is converted to a colored complex and absorbance is read at 680â¯nm; similarly, urea concentration in blood and urine is determined via enzymatic color reactions and color intensity is quantified by colorimetric measurement.